Development of a chemiluminescent immunoassay based on magnetic solid phase for quantification of homocysteine in human serum

Abstract

Background

Homocysteine (HCY) is a sulfur-containing amino acid that is an independent or important risk factor for the occurrence of many chronic diseases and is one of the most important indicators for determining health risks. However, existing HCY detection methods do not meet the requirements of clinical diagnosis. Therefore, there is an urgent need to establish new detection methods to meet the needs of clinical detection.

Results

In this study, we used the principle of competitive method to establish a new method for the determination of HCY in human serum using a chemiluminescent enzyme immunoassay in conjunction with a chemiluminescent assay instrument that uses magnetic microparticles as the solid phase of the immunoreaction. The established method achieved satisfactory results in terms of minimum detection limit, specificity, accuracy, and clinical application. The limit of detection was 0.03 ng/mL. The intra-assay coefficient of variation (CV) was 1.94–5.05%, the inter-assay CV was 2.29–6.88%, and the recovery rate was 88.60–93.27%. Cross-reactivity with L-cysteine ranged from 0.0100 to 0.0200 μmol/L, and cross-reactivity with glutathione ranged from 0.0100 to 0.200 μmol/L, all of which were less than the limit of detection (LoD) of this method. The linear factor R of this method was greater than 0.99.

Conclusions

In summary, the developed method showed a good correlation with the product from Abbott. A total of 996 clinical patients with cardiovascular diseases were evaluated using the method developed in this study.

Peer Review reports

Background

The human body cannot directly ingest or synthesize homocysteine (HCY) [1]. HCY is an intermediate product produced during the metabolism of methionine and cysteine [2]. Methionine, an essential amino acid derived from dietary proteins such as poultry, meat, eggs, seafood, and dairy products, is an exclusive dietary source of HCY [3]. The liver plays a pivotal role in regulating HCY levels, as it produces enzymes essential for methionine metabolism [4, 5]. Methionine is initially transformed by methionine adenosyltransferase into S-adenosylmethionine (SAM) [6], a highly energetic compound, with adenosine triphosphate supplying the adenosyl group. Subsequently, SAM donates a methyl group via methyltransferase to form S-adenosylhomocysteine (SAH) [6]. Finally, SAH hydrolase catalyzes the conversion of SAH to HCY [6]. Therefore, the equilibrium of products and metabolic balance of HCY is crucial for maintaining human homeostasis.

Excessive or insufficient HCY concentrations have been consistently shown to negatively affect human health [3, 7]. Recent research findings have indicated a statistically significant correlation between elevated HCY levels and a range of health issues [7], including disorders in the cardiovascular [8], gastrointestinal [9], urinary [10], skeletal [11], and autoimmune systems [12], as well as cancer [13]. Several methods have been developed to detect serum HCY levels in clinical settings [14], including gas chromatography-mass spectrometry [15], liquid chromatography-mass spectrometry [16], radioenzyme analysis [17], capillary electrophoresis [18], high-performance liquid chromatography (HPLC)- fluorescence detection [19], enzyme-linked immunosorbent assays [20], chemiluminescent microparticle immunoassays [21], fluorescence polarization immunoassays [22], and HPLC [23]. Nevertheless, no universally accepted method of detection exists because these methods are usually limited by expensive instruments and detection reagents or insufficient sensitivity [3]. Therefore, a method that can rapidly and accurately detect HCY based on simple operation is urgently needed.

Since magnetic particle-based chemiluminescence immunoassay is based on a liquid-phase reaction that allows for sufficient contact between the substrate and reactants, it has clear advantages in terms of linear range, sensitivity, precision, accuracy, and good detection stability [24,25,26,27,28,29]. Furthermore, detection is rapid and convenient with fully automated instruments. Given these advantages, a sandwich chemiluminescence assay was established to detect HCY levels in human serum.

Methods

Reagents and materials

The anti-SAH-AP conjugate is a conjugation reaction between the SAH monoclonal antibody and alkaline phosphatase (AP), purified to form a conjugation, dithiothreitol (DTT), and recombinant S-adenosyl-L homocysteine hydrolase (SAHase), obtained from ZECEN Biotech Co., Ltd (Taizhou, Jiangsu Province, China). Magnetic particle conjugates formed by linking SAH antibodies to magnetic particles were supplied by ZECEN Biotech Co., Ltd. (Taizhou, Jiangsu Province, China). The HCY recombinant antigen was supplied by ZECEN Biotech Co., Ltd. (Taizhou, Jiangsu Province, China). BSA was purchased from ROCHE (Basel, Switzerland). Carboxyl magnetic beads were purchased from BIOMAG (Wuxi, Jiangsu Province, China). The manufacturer and model of the chemiluminescence analyzer utilized was the CIA1200 Full-auto Chemiluminescence Analyzer (ZECEN Biotech Co., Ltd, Taizhou, Jiangsu Province, China).

Reagent component preparation

Characterization of anti-SAH-AP conjugate, SAHsae, and reducing agent

Anti-SAH-AP (2 μg/mL): add SAH monoclonal antibody conjugate (SAH antibody alkaline phosphatase AP) to buffer containing BSA (pH = 8.0) to prepare SAH antibody reagent. Reducing agent (2 μg/mL): prepared by adding dithiothreitol (DTT) to the citrate buffer solution. SAHase (2 μg/mL): recombinant S-adenosyl-L-hydrocysteine hydrolase (SAHase) was prepared by adding it to tris containing buffer. Magnetic particle reagent (3 mg/mL): magnetic particles and SAH connectors in BSA buffer (pH = 8.0). The concentration and recovery rate of the antibodies were calculated by measuring their absorbance values at 280 nm using a UV spectrophotometer. The selected concentration of anti-SAH-AP was 2.0 mg/mL, the reducing agent (DTT) concentration was 3. 0 mg/mL, and the SAHase concentration was 3.0 mg/mL. The recovery rates were 100.70%, 102.90%, and 105.13%, respectively.

Reaction condition optimization

This reagent kit adopts the principle of competition method, using magnetic microparticles as the solid phase of the immune reaction, in addition to using chemiluminescence enzyme-linked immunosorbent assay method in combination with chemiluminescence measuring instruments to determine the HCY concentration in human serum.

The technical principle is that bound or dimerized HCY (oxidized) is reduced to free HCY by the action of dithiothreitol (DTT). In the presence of sufficient adenosine, free HCY is converted to S-adenosine homocysteine (SAH) by recombinant S-adenosine homocysteine hydrolase (rSAHHase). SAH competes with magnetic particle-labeled SAH to bind to alkaline phosphatase (AP)-labeled anti-SAH monoclonal antibodies. It directly precipitates in an external magnetic field to separate the complexes formed by immune reactions from other unbound substances. The precipitated complex was cleaned by adding an enzyme-catalyzed chemiluminescent substrate, which is catalytic cracked under the action of the enzyme, forming an unstable excited state intermediate. When the excited state intermediate returns to the ground state, it emits photons, forming a luminescent reaction. The luminescence intensity of the reaction is measured using a luminescent instrument. Within the measurement range, luminescence intensity is inversely proportional to the concentration of HCY in the sample. The improved Four-Parameter Logistic (4PL) Model can quantitatively calculate the HCY concentration in the test sample. In this study, factors such as magnetic particle, anti-SAH-AP, and DTT concentrations and incubation time were optimized.

For the details of the 4PL model, it is defined by the following equation:

$$y = \frac{{{A_1} - {A_0}}}{{1 + {{\left( {\frac{x}{{{x_0}}}} \right)}^p}}} + {A_0}$$

Curve Shape: S-shaped, either increasing or decreasing.

A1: The maximum value of y as x approaches infinity or negative infinity.

A0: The minimum value of y as x approaches infinity or negative infinity.

X: The inflection point of the curve.

P: Related to the slope of the curve at the inflection point.

Based on the above formula, we will substitute the following data for calculation.

Concentration of Standard Sample

RLUs

0 μmol/L

1,137,651

2.5 μmol/L

983,113

5 μmol/L

852,651

10 μmol/L

686,961

20 μmol/L

466,081

50 μmol/L

313,807

We obtain the following curve:

figure a

And obtain the following formula for the concentration calculation of subsequent test samples:

$$\begin{array}{l}{\rm{Y = 1135118}}{\rm{.3913 + }}\left( {{\rm{ - 953720}}{\rm{.5979}}} \right){\rm{/}}\\{\rm{(1 + EXP}}\left( {{\rm{ - }}\left( {{\rm{ - 2}}{\rm{.8019 + 1}}{\rm{.1945*Ln}}\left( {\rm{X}} \right)} \right)} \right)\end{array}$$

Method procedure

First, add 30 μL of calibration, test samples, or quality control products into the reaction, then add 15 μL DTT, mix well, and react at 37 ℃ for 5 min. Then, add 15 μL SAHase into the reaction and mix well; add 30 μL magnetic particle reagent and 30 μL anti-SAH-AP, mix, and react at 37 ℃ for 30 min. Then, add 300 μL of cleaning solution for the chemiluminescence system and mix well. Repeat this step three times for magnetic separation and cleaning of the reactants. Finally, add 200 μL chemiluminescent substrate solution for automatic immunoassay for luminescence value detection, and concentration values are calculated if the calibrator curve is available.

Sample collection

The selection criteria are as follows: A specific proportion of samples exhibiting aberrant values outside the established reference range and a corresponding proportion of samples displaying normal values are to be identified, with a minimum total sample size of 120 and a minimum of 30% of the total number of samples exhibiting aberrant values (a minimum of 36 samples); this is to be achieved by utilizing the control product. Samples with abnormal values are expected to display a range of HCY levels. All selected samples must meet the requisite information for this validation process. It is imperative that specimens exhibiting severe hyperlipidemia, hemolysis, or jaundice be excluded from the study. Specimens that do not align with the clinical situation, as indicated by the comparative reagent assay results, are excluded from further analysis.

To establish the reference interval for this kit, 800 serum specimens from healthy (or normal) individuals were collected from a health-testing center within our hospital. Of these, 400 were from individuals aged over 70 years, and 400 were from individuals aged under 70 years. The 95th percentile of the test values for the normal human samples was calculated as the reference value, with a lower limit of 2.5% and an upper limit of 97.5%.

Performance test method

Limit of detection (LoD)

Using zero concentration calibration samples or sample diluents for detection, repeat the measurement 20 times to obtain the relative luminescence value (RLU) of the 20 measurement results. Calculate the average value (M) and standard deviation (SD) to obtain M-2SD. Based on the concentration RLU value results between point A calibration samples and adjacent point B calibration samples, perform a two-point regression fitting to obtain a first-order equation. Substitute the RLU value of M-2SD into the above equation and calculate the corresponding concentration value, which is the lowest detection limit.

Precision and recovery

Take quality control samples 1 and 2 from the same batch of reagent kits as the test samples. Each test is repeated ten times, with a coefficient of variation (CV) not exceeding 8% for the intra-assay. Using three different batches of reagent kits and following the instructions, use a chemiluminescence analyzer to measure the samples separately 30 times (10 times per batch of reagent kits). The CV between the three batches of reagent kits should not exceed 15%. Add 10 μL 50 μmol/L sample A to 90 μL 2.5 μmol/L sample B, and the volume of A added should not exceed 10% of the total volume (A + B), and the recovery rate was calculated.

Accelerated stability

After determining all the components in the reagent kit, prepare small samples and divide them into five portions (100 tests). Place them in a 37 ℃ incubator for 1, 2, 3, and 7 days, and measure the performance of the reagent kit at 37 ℃ for 0 h, 1, 2, 3, and 7 days, respectively. Conduct performance tracking experiments on three batches of reagent kits under long-term actual storage conditions (2 ℃ to 8 ℃). Determine the shelf life of the three batches of reagent kits based on their stability under storage conditions. The appearance and performance indicators (linearity, accuracy, minimum detection limit, and repeatability) of the test kit are tracked and tested by placing it under conditions of 2 ℃ to 8 ℃ for 10 months and then inspecting it once every 3, 6, 9, and 10 months.

Linearity

Dilute high-value samples close to the upper limit of the linear range to at least five concentrations in a certain proportion, with low-value samples approaching the lower limit of the linear range. According to the instructions of the reagent kit, repeat the test three times for each concentration of the sample, calculate the average value, fit the average value and dilution ratio using the least squares method, and calculate the linear correlation coefficient r. The linear range of the HCY reagent kit is 0.4–50.0 μmol/L. Within this linear range, the correlation coefficient r between the reference and test curve should be less than 0.9900.

Method comparison

The established method was compared with an on-market CLIA method from Abbott. The sample size used to compare with Abbott was 240 and ranged from 4.04 μmol/L to 40.98 μmol/L.

Results

Serum HCY chemiluminescence assay development

This method adopts the principle of competitive method, using magnetic microparticles as the solid phase of the immunoreaction and the chemiluminescent enzyme immunoassay method with a chemiluminescent measuring instrument to determine the level of free HCY in human serum. A stable and reliable experimental procedure was determined by optimizing the SAH antibody coupler, reductant DTT, and SAH hydrolase involved in the method and systematically optimizing the sample volume, reaction time, and temperature (Fig. 1).

Fig. 1
figure 1

Scheme of the applied procedure

Method performance

The 4-parameter logic function method was used to fit the standard curve of the signal with HCY concentration. The following series of concentrations of HCY were used as a standard curve (A-F): 0, 2.5, 5.0, 10.0, 20.0, and 50.0 μmol/L. The LoD is 0.3 μmol/L. Notably, the coefficient of variation (CV) is a statistic employed to ascertain the extent of dispersion of the constituent data points a dataset around its mean; this is a valuable metric in the context of comparative analysis as it allows for the assessment of variation across data series, even when there is considerable divergence in their respective means. Therefore, the precision was tested; the intra-assay CV % was 1.94 – 5.05%, and the Inter-assay CV % was 2.29 – 6.88%; the results are shown in Table 1. Considering that the kit is required to demonstrate an intra-assay CV of less than 8% and an inter-assay CV of less than 15%, the data indicate that the developed HCY kit has satisfactory accuracy and reproducibility, highlighting its commendable analytical performance and capacity to meet the demands of clinical practice.

Table 1 Precision test results

Advice on developed method

Optimization of the sample volume and antibody concentration in the reagent

As shown in Table 2, a series of concentrations of anti-SAH-AP and DTT were tested. Additional free antibodies were added to the reaction system to consume excess antigens in the samples. As shown in Fig. 2, the greatest discrepancy in absorbance values was observed between the three standards when a sample volume of 30 μl was employed. Therefore, the sample volume was optimized at 30 μl. Additionally, the DTT, SAHase, and anti-SAH-AP reaction dosages were optimized (These components were explored by taking 15 μl at indicated concentrations in the table). As listed in Table 2, at a dose of 2.0 μg/mL, the difference in absorbance values between the three standards was the most significant, and the linearity was optimal (The units of numbers in the table are μg/mL). Finally, 30 μL anti-SAH-AP (2.0 μg/mL), 15 μL DTT (2.0 μg/mL), and 15 μL SAHase (2.0 μg/mL) were selected as the optimal formation, as shown in Table 3.

Table 2 Optimization of the antibody concentration
Fig. 2
figure 2

Optimization of sample volume

Table 3 Optimization of the HCY anti-reagent volume

Optimization of the incubation time

The anti-SAH-AP and sample incubation times were assessed. In addition to the incubation time, the procedure was performed as described in the Methods. The RLUs increased with increasing incubation time and remained unchanged after 30 min when the reaction reached dynamic equilibrium. Owing to time-saving clinical requirements, we selected a 30 min incubation time. The signal results are shown in Fig. 3.

Fig. 3
figure 3

Optimization of incubation time

Reaction flow and reaction parameters

To this end, the aforementioned data illustrates that the following reaction flow applies to this kit: the user is required to select the corresponding test item on the instrument’s operation interface, generate the work list, select ‘start testing,’ and then allow the instrument to perform automatic testing in accordance with the item’s set reaction flow and reaction parameters. The details are presented in Table 4.

Table 4 Reaction flow and reaction parameters

Comparison of methods

In this study, 240 serum samples were analyzed using both the developed method and the Abbott method, which is a clinically utilized detection method for HCY in the context of cardiovascular disease diagnosis [21]. The test results were regressed using the least-squares regression equation, and the correlation coefficient was computed. Data are shown in Fig. 4. The test results show good agreement between the developed and compared methods, with a correlation coefficient (R2) of 0.95, and the difference in the test and mean values indicated a slight bias between these methods. Notably, the discrepancies in HCY values observed between the two methods were predominantly evident in samples from patients with elevated HCY levels. Based on our experience in the development of the kits, we postulate that the potential causes for this may include epitopes recognized by the antibodies, as well as elevated levels of other substances that may interfere with the detection of HCY.

Fig. 4
figure 4

Comparison of the developed method and Abbott method

Clinical study

The most crucial element in evaluating the efficacy of the kits we create is their precision in a clinical context. First, to determine the reference interval of the kit, 800 serum samples were collected from healthy individuals at our hospital’s Health and Wellness Center. Of these, 400 were from individuals aged over 70 years, and 400 were from individuals under 70 years old. Furthermore, a bilateral 95% reference value was calculated using the assay values of normal human samples, with a lower limit of 2.5% and an upper limit of 97.5%. The final reference values for this kit were determined as ≤ 70 years: 4–14 μmol/L; >70 years: 6–20 μmol/L. Given that an elevated HCY level has been identified as a key biomarker for cardiovascular disease [8, 30], we collected serum samples from 996 patients with cardiovascular disease to evaluate the precision of our kit. As shown in Table 5, the sensitivity of our kit for the detection of disease was 93.4% (930 996), suggesting that our kits have the potential for clinical use. Consistent with a previous study [31], we also observed that HCY levels in men were slightly higher than those in the serum collected from women (Fig. 5).

Table 5 Distribution of clinical samples
Fig. 5
figure 5

HCY concentration in women and men

Conclusions

The forms of HCY present in plasma include reduced HCY, disulfide HCY, and mixed HCY [32]. Under normal conditions, the plasma contains very little free HCY, mainly in the form of protein [33]. Many factors, such as genetics, nutrition, or drugs, can affect the metabolism of cysteine in the body, resulting in hyperhomocysteinemia [34]. High levels of HCY are associated with many diseases [7], such as Alzheimer’s disease [35], diabetes [36], cancer [37], neurological disorders [38], age-related macular degeneration [39], vascular disease [40], and hearing loss [41]. Commonly used methods for HCY detection include HPLC, LC-MS, fluorescence polarization immunoassay (FPLA), enzyme immunoassay (ELISA), and cyclic enzyme assays [14,15,16,17,18,19,20,21,22,23]. However, these methods have disadvantages, such as the need for specialized equipment and technicians and a long operation time, which cannot meet the needs of the clinical testing of individual and emergency samples [3]. Of note, magnetic particle-based chemiluminescence immunoassay represents a novel analytical technique that integrates magnetic separation, chemiluminescence, and immunoassay [24,25,26,27,28,29]. This technology optimally exploits the rapid and automated magnetic separation, high sensitivity of chemiluminescence technology, and specificity of immunoassays [24,25,26,27,28,29]. These characteristics contribute to the high specificity, sensitivity, speed, and simplicity of CLIA, which has consequently been widely adopted in disease diagnosis [24,25,26,27,28,29]. Therefore, this kit adopts advanced magnetic separation chemiluminescence immunoassay technology to detect HCY in blood. Through research on the process and reaction system, we determined the source of raw materials as well as quality standards and testing methods, formulated the production process, operation flow, quality standards and testing methods for each component, and established product standards and corresponding testing methods for the finished kit. In addition, we carried out a trial production of the product according to the established production process and quality control and evaluated the performance of the kit according to the determined reaction system and operation. The results showed that the production process of this product is stable, its quality is controllable, the reaction system is reasonable, and it meets the relevant technical, production, and clinical requirements. The results of clinical tests using this kit showed a good correlation with diseases and met clinical requirements. However, if this kit is to be used clinically in the future, a larger number of clinical test samples will be required for validation. Minimizing the variation in raw material lot helps maintain the performance of the kit.

Of note, serum HCY levels can be influenced by various factors, including age [42], sex [42], medication [28], and other variables [29]. The intake of methionine and the supply of folate and vitamins B6 and B12 are among the dietary factors that influence HCY levels [43]. Genetic defects, polymorphisms, or mutations in several key enzymes can impair HCY metabolism, leading to elevated HCY levels [29]. Additionally, some studies have observed a sex difference in plasma HCY levels [42]. Furthermore, HCY levels are higher in individuals aged 55 and above than in those below 55 years of age [44]. Based on these findings, we propose that the HCY reference value should be contextualized in clinical applications.

Data availability

Data is provided within the manuscript information files. And, all other data and materials supporting the findings of this study are available from the corresponding authors upon reasonable request.

Abbreviations

HCY:

Homocysteine

HPLC:

High Performance Liquid Chromatography

LC-MS:

Liquid Chromatograph Mass Spectrometer

LoD:

Limit of Detection

References

  1. Jakubowski H. Translational incorporation of S-nitrosohomocysteine into protein. J Biol Chem. 2000;275(29):21813–6.

    Article  CAS  PubMed  Google Scholar 

  2. Jakubowski H. Homocysteine editing, thioester chemistry, coenzyme A, and the origin of coded peptide synthesis dagger. Life (Basel). 2017;7(1).

  3. Wu DF, Yin RX, Deng JL. Homocysteine, hyperhomocysteinemia, and H-type hypertension. Eur J Prev Cardiol. 2024;31(9):1092–103.

    Article  PubMed  Google Scholar 

  4. Teng YW, Mehedint MG, Garrow TA, Zeisel SH. Deletion of betaine-homocysteine S-methyltransferase in mice perturbs choline and 1-carbon metabolism, resulting in fatty liver and hepatocellular carcinomas. J Biol Chem. 2011;286(42):36258–67.

    Article  CAS  PubMed  Google Scholar 

  5. Stender S, Chakrabarti RS, Xing C, Gotway G, Cohen JC, Hobbs HH. Adult-onset liver disease and hepatocellular carcinoma in S-adenosylhomocysteine hydrolase deficiency. Mol Genet Metab. 2015;116(4):269–74.

    Article  CAS  PubMed  Google Scholar 

  6. Selhub J. Homocysteine metabolism. Annu Rev Nutr. 1999;19:217–46.

    Article  CAS  PubMed  Google Scholar 

  7. Smith AD, Refsum H. Homocysteine - from disease biomarker to disease prevention. J Intern Med. 2021;290(4):826–54.

    Article  CAS  PubMed  Google Scholar 

  8. Kozakova M, Morizzo C, Penno G, Shore AC, Nilsson J, Palombo C. Plasma homocysteine and cardiovascular organ damage in a population with a high prevalence of risk factors. J Clin Endocrinol Metab. 2020;105(8).

  9. Givvimani S, Munjal C, Narayanan N, Aqil F, Tyagi G, Metreveli N, Tyagi SC. Hyperhomocysteinemia decreases intestinal motility leading to constipation. Am J Physiol Gastrointest Liver Physiol. 2012;303(3):G281–290.

    Article  CAS  PubMed  Google Scholar 

  10. Friedman AN, Bostom AG, Selhub J, Levey AS, Rosenberg IH. The kidney and homocysteine metabolism. J Am Soc Nephrol. 2001;12(10):2181–9.

    Article  CAS  PubMed  Google Scholar 

  11. Ungvari Z, Pacher P, Rischak K, Szollar L, Koller A. Dysfunction of nitric oxide mediation in isolated rat arterioles with methionine diet-induced hyperhomocysteinemia. Arterioscler Thromb Vasc Biol. 1999;19(8):1899–904.

    Article  CAS  PubMed  Google Scholar 

  12. Lazzerini PE, Capecchi PL, Selvi E, Lorenzini S, Bisogno S, Galeazzi M. Laghi Pasini F: hyperhomocysteinemia, inflammation and autoimmunity. Autoimmun Rev. 2007;6(7):503–9.

    Article  CAS  PubMed  Google Scholar 

  13. Xie H, Wei L, Wang Q, Tang S, Gan J. Elevated serum homocysteine levels associated with poor recurrence-free and overall survival in patients with colorectal cancer. Sci Rep. 2024;14(1):10057.

    Article  CAS  PubMed  Google Scholar 

  14. Ryan A, Twomey PJ. Homocystinuria: a commentary. J Clin Pathol. 2023;76(3):153–5.

    Article  PubMed  Google Scholar 

  15. Nelson BC, Pfeiffer CM, Zhang M, Duewer DL, Sharpless KE, Lippa KA. Commutability of NIST SRM 1955 homocysteine and folate in frozen human serum with selected total homocysteine immunoassays and enzymatic assays. Clin Chim Acta. 2008;395(1–2):99–105.

    Article  CAS  PubMed  Google Scholar 

  16. Gopu CL, Hari PR, George R, Harikrishnan S, Sreenivasan K. Simultaneous determination of homocysteine and asymmetric dimethylarginine in human urine by liquid chromatography-tandem mass spectrometry. J Chromatogr B Analyt Technol Biomed Life Sci. 2013;939:32–7.

    Article  CAS  PubMed  Google Scholar 

  17. Park SH, Lee JY, Cho HN, Kim KR, Yang SA, Kim HJ, Jhee KH. Simple and novel assay of the host-guest complexation of homocysteine with Cucurbit[7]uril. J Microbiol Biotechnol. 2019;29(1):114–26.

    Article  CAS  PubMed  Google Scholar 

  18. Causse E, Siri N, Bellet H, Champagne S, Bayle C, Valdiguie P, Salvayre R, Couderc F. Plasma homocysteine determined by capillary electrophoresis with laser-induced fluorescence detection. Clin Chem. 1999;45(3):412–4.

    Article  CAS  PubMed  Google Scholar 

  19. Vester B, Rasmussen K. High performance liquid chromatography method for rapid and accurate determination of homocysteine in plasma and serum. Eur J Clin Chem Clin Biochem. 1991;29(9):549–54.

    CAS  PubMed  Google Scholar 

  20. Quintana I, Freeman D, Galarza C, Murua A, Spence JD, Kordich L. Validation of an enzyme immunoassay for the determination of total homocysteine in plasma. Blood Coagul Fibrinolysis. 2000;11(3):235–8.

    CAS  PubMed  Google Scholar 

  21. Kim HN, Yoon SY. Comparison of a new enzymatic assay for serum homocysteine on Toshiba TBA-c16000 against an immunoassay on Abbott Architect. Scand J Clin Lab Invest. 2021;81(1):46–51.

    Article  CAS  PubMed  Google Scholar 

  22. Nexo E, Engbaek F, Ueland PM, Westby C, O’Gorman P, Johnston C, Kase BF, Guttormsen AB, Alfheim I, McPartlin J, et al. Evaluation of novel assays in clinical chemistry: quantification of plasma total homocysteine. Clin Chem. 2000;46(8 Pt 1):1150–6.

    Article  CAS  PubMed  Google Scholar 

  23. Pfeiffer CM, Huff DL, Gunter EW. Rapid and accurate HPLC assay for plasma total homocysteine and cysteine in a clinical laboratory setting. Clin Chem. 1999;45(2):290–2.

    Article  CAS  PubMed  Google Scholar 

  24. Matsunaga T, Kawasaki M, Yu X, Tsujimura N, Nakamura N. Chemiluminescence enzyme immunoassay using bacterial magnetic particles. Anal Chem. 1996;68(20):3551–4.

    Article  CAS  PubMed  Google Scholar 

  25. Fu X, Li L, Wu G, Tang K, Zhang J, Chen Z, Shi M, Zhang B. Establishment of sensitive sandwich-type chemiluminescence immunoassay for interleukin-18 in urinary samples. Appl Biochem Biotechnol. 2023;195(12):7414–28.

    Article  CAS  PubMed  Google Scholar 

  26. Liao M, Zheng J, Xu Y, Qiu Y, Xia C, Zhong Z, Liu L, Liu H, Liu R, Liang S. Development of magnetic particle-based chemiluminescence immunoassay for measurement of human procalcitonin in serum. J Immunol Methods. 2021;488:112913.

    Article  CAS  PubMed  Google Scholar 

  27. Chen C, Liang H, Peng F, Zhong S, Lu Y, Guo G, Li L. Determination of echinococcosis IgG antibodies using magnetic bead-based chemiluminescence immunoassay. J Immunol Methods. 2023;520:113513.

    Article  CAS  PubMed  Google Scholar 

  28. Kim J, Kim H, Roh H, Kwon Y. Causes of hyperhomocysteinemia and its pathological significance. Arch Pharm Res. 2018;41(4):372–83.

    Article  CAS  PubMed  Google Scholar 

  29. Liao S, Guo S, Ma R, He J, Yan Y, Zhang X, Wang X, Cao B, Guo H. Association between methylenetetrahydrofolate reductase (MTHFR) C677T polymorphism and H-type hypertension: a systematic review and meta-analysis. Ann Hum Genet. 2022;86(5):278–89.

    Article  CAS  PubMed  Google Scholar 

  30. Koklesova L, Mazurakova A, Samec M, Biringer K, Samuel SM, Busselberg D, Kubatka P, Golubnitschaja O. Homocysteine metabolism as the target for predictive medical approach, disease prevention, prognosis, and treatments tailored to the person. EPMA J. 2021;12(4):477–505.

    Article  PubMed  Google Scholar 

  31. Onyemelukwe OU, Maiha BB. Prevalence of hyperhomocysteinaemia, selected determinants and relation to hypertension severity in Northern-nigerian hypertensives: the ABU homocysteine survey. Ghana Med J. 2020;54(1):17–29.

    Article  PubMed  Google Scholar 

  32. Jakubowski H. Homocysteine thiolactone: metabolic origin and protein homocysteinylation in humans. J Nutr. 2000;130(2S Suppl):s377–81.

    Article  Google Scholar 

  33. Hermann A, Sitdikova G. Homocysteine: biochemistry, molecular biology and role in disease. Biomolecules. 2021;11(5).

  34. Sleeman I, Lawson RA, Yarnall AJ, Duncan GW, Johnston F, Khoo TK, Burn DJ. Urate and homocysteine: predicting motor and cognitive changes in newly diagnosed Parkinson’s disease. J Parkinsons Dis. 2019;9(2):351–9.

    Article  CAS  PubMed  Google Scholar 

  35. Sade Yazdi D, Laor Bar-Yosef D, Adsi H, Kreiser T, Sigal S, Bera S, Zaguri D, Shaham-Niv S, Oluwatoba DS, Levy D et al. Homocysteine fibrillar assemblies display cross-talk with Alzheimer’s disease β-amyloid polypeptide. Proc Natl Acad Sci U S A. 2021;118(24).

  36. Mursleen MT, Riaz S. Implication of homocysteine in diabetes and impact of folate and vitamin B12 in diabetic population. Diabetes Metab Syndr. 2017;11(Suppl 1):S141–6.

    Article  PubMed  Google Scholar 

  37. Zhang D, Wen X, Wu W, Guo Y, Cui W. Elevated homocysteine level and folate deficiency associated with increased overall risk of carcinogenesis: meta-analysis of 83 case-control studies involving 35,758 individuals. PLoS ONE. 2015;10(5):e0123423.

    Article  PubMed  Google Scholar 

  38. Diaz-Arrastia R. Homocysteine and neurologic disease. Arch Neurol. 2000;57(10):1422–7.

    Article  CAS  PubMed  Google Scholar 

  39. Pinna A, Zaccheddu F, Boscia F, Carru C, Solinas G. Homocysteine and risk of age-related macular degeneration: a systematic review and meta-analysis. Acta Ophthalmol. 2018;96(3):e269–76.

    Article  CAS  PubMed  Google Scholar 

  40. Hankey GJ, Eikelboom JW. Homocysteine and vascular disease. Lancet. 1999;354(9176):407–13.

    Article  CAS  PubMed  Google Scholar 

  41. Martínez-Vega R, Garrido F, Partearroyo T, Cediel R, Zeisel SH, Martínez-Álvarez C, Varela-Moreiras G, Varela-Nieto I, Pajares MA. Folic acid deficiency induces premature hearing loss through mechanisms involving cochlear oxidative stress and impairment of homocysteine metabolism. Faseb j. 2015;29(2):418–32.

    Article  PubMed  Google Scholar 

  42. Xu R, Huang F, Wang Y, Liu Q, Lv Y, Zhang Q. Gender- and age-related differences in homocysteine concentration: a cross-sectional study of the general population of China. Sci Rep. 2020;10(1):17401.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  43. Song W, Qin Z, Hu X, Han H, Li A, Zhou X, Li Y, Li R. Using Bayesian networks with tabu-search algorithm to explore risk factors for hyperhomocysteinemia. Sci Rep. 2023;13(1):1610.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  44. Chori BS, Danladi B, Inyang BA, Okoh MP, Nwegbu MM, Alli AL, Odili AN. Hyperhomocysteinemia and its relations to conventional risk factors for cardiovascular diseases in adult nigerians: the REMAH study. BMC Cardiovasc Disord. 2021;21(1):102.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank the support from Wuxi Translational Medicine Research Institution for this research.

Funding

This work was supported by the Project of Wuxi Translational Medicine Research Institution (Project number: LCYJ202309).

Author information

Authors and Affiliations

Authors

Contributions

This project was conceived and supervised by YKL and XWQ. The method was designed by NS and XWQ (Fig. 1). Method performance validation, optimization (Figs. 2, 3 and 4), and all corresponding statistical analyses were performed by YKL and YRL. HCY concentrations of different patients (Fig. 5) were determined by RW. Patient sample collection was contributed by SNC. YKL, NS, and XWQ prepared the manuscript. All the authors reviewed the manuscript.

Corresponding authors

Correspondence to Ning Sun or Xiaowei Qi.

Ethics declarations

Ethics approval

This study was conducted in accordance with the Declaration of Helsinki. Prior to obtaining patient samples, requisite approval from the Affiliated Hospital of Jiangnan University and written informed consent were obtained from all patients.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Liu, Y., Li, Y., Wang, R. et al. Development of a chemiluminescent immunoassay based on magnetic solid phase for quantification of homocysteine in human serum. BMC Biotechnol 24, 77 (2024). https://doi.org/10.1186/s12896-024-00903-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12896-024-00903-z

Keywords