Table 1 Calibration results for each compound

From: Does size matter? Separations on guard columns for fast sample analysis applied to bioenergy research

Compound

Detection

Linear quantification range

Regression coefficient

Ethanol

RID

0.01 − 50 mg/mL

0.9995

Acetone

RID

0.01 − 50 mg/mL

0.9996

 

UV 265 nm

0.01 − 50 mg/mL

0.9999

 

UV 285 nm

0.02 − 50 mg/mL

0.9995

iso-Butanol

RID

0.005 − 50 mg/mL

0.9999

n-Butanol

RID

0.005 − 50 mg/mL

0.9997

5-HMF

RID

0.005 − 50 mg/mL

0.9999

 

UV 280 nm

0.00005 − 0.25 mg/mL

0.9999

Furfural

RID

0.005 − 50 mg/mL

0.9995

 

UV 280 nm

0.00005 − 0.25 mg/mL

0.9999

Glucose

RID

0.005 − 20 mg/mL

0.9999

Xylose

RID

0.005 − 20 mg/mL

0.9999

  1. Chromatography conditions: 50 mm × 7.8 mm RFQ Fast Acid H+ kept at 30 °C, flow rate: 1.0 ml/min of 5 mM sulfuric acid, 20 μL injection volume; Lower limit of the calibration range was determined as the concentration level with a relative standard deviation (RSD) <10 % (n = 5)