Fig. 11 | BMC Biotechnology

Fig. 11

From: Validation and calibration of a novel GEM biosensor for specific detection of Cd2+, Zn2+, and Pb2+

Fig. 11

Calibration plot of individual heavy metal treatments between fluorescent intensities in CTCF and heavy metal concentrations in ppb. The fluorescent intensities have been obtained for samples containing metal ion concentrations in between 1 to 6 ppb. The treatments with Ni2+, Pb2+, Zn2+ and Cd2+ are shown in black colour bars, orange colour bars, blue colour bars and red colour bars in order. The R2 values for Ni2+, Pb2+, Zn2+ and Cd2+ trends are as 0.3825, 0.9785, 0.9761 and 0.9809. The error bars represent standard mean differences of three independent biological treatments. A single asterisk (*) represents a P value less than 0.05 (P < 0.05), indicating statistical significance at the 0.05 level. Double asterisks (**) represent a P value less than 0.01 (P < 0.01), indicating statistical significance at the 0.01 level. Triple asterisks (***) represent a P value less than 0.001 (P < 0.001), indicating statistical significance at the 0.001 level. Quadruple asterisks (****) represent a P value less than 0.0001 (P < 0.0001), indicating statistical significance at the 0.0001 level