Table 1 Determination of accuracy of the RSSAA method

From: A high throughput method for total alcohol determination in fermentation broths

NO. Ethanol concentration spiked Ethanol concentration detecteda Detection accuracy
  (%,V/V) (%,V/V) (%)
1 1.93 2.01 ± 0.09 104.1 ± 4.7
2 3.79 3.85 ± 0.12 101.6 ± 3.2
3 5.57 5.98 ± 0.28 107.4 ± 5
4 7.3 7.32 ± 0.15 100.3 ± 2.1
5 8.95 9.08 ± 0.09 101.5 ± 1
6 10.55 10.61 ± 0.11 100.6 ± 1
7 12.11 12 ± 0.12 99.1 ± 1
8 13.59 13.72 ± 0.25 101 ± 1.8
9 15.02 15.12 ± 0.21 100.7 ± 1.4
10 16.42 16.44 ± 0.02 100.1 ± 0.1
  1. aThe values = measured values - 1.5% (ethanol concentration in PS sample)